Electro-control: am EMG-controlled A/K prosthesis

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

Locking of the flexion of the knee-joint in this knee prosthesis can be accomplished by means of a mechanical clutch released by a control pulse which, in turn, is derived from the myolectric activity of the patient's stump. The feature enables the amputee to use the artificial leg as a reaction point for his residual muscles.

Ascending stairs and ramps using the prosthesis is made possible by the use of a mechanical knee-joint which, when locked, cannot be flexed any more, but is still free to extend under the patient's control.

The prosthesis described in this paper employs a rather unconventional control system. Pulse triggering, instead of pulse integration, is used. A logic device inhibits the detection system once the knee lock has occurred; the whole system is automatically re-set as soon as the patient fully extends the artificial leg.

A patient can be trained to use the leg for climibing stairs in a matter of hours.

Sommaire

Le blocage de la flexion de l'articulation du genou dans cette prothèse peut être accompli à l'aide d'un embrayage mécanique relâché par un pouls de contrôle qui à son tour est dérivé d'une activité myoélectrique du moignon du patient. Cette caractéristique permet à l'amputé d'user la jambe artificielle comme un point de réaction pour ses muscles résiduels.

Il devient possible de monter les escaliers et les rampes en utilisant la prothèse, à l'aide d'une articulation mécanique qui, quand elle est bloquée, ne peut plus ètre fléchie, mais qui est néanmoins libre de s'étendre sous le contrôle du patient.

La prothèse décrite dans ce mémoire utilise un système de contrôle plutôt original. On utilise une détente par puls au lieu de l'intégration du pouls. Un dispositif logique inhibe le système de détection, du moment que le blocage du genous se produit; le systèmme entier est automatiquement remis en fonction aussitôt que le patient étend complèment la jambe artificielle.

Un patient peut être instruit en quelques heures d'utiliser la jambe pour monter des marches.

Zusammenfassung

Sperrung der Flexion des Kniegelenkes in dieser Knieprothese kann mittels einer mechanischen Kupplung erreicht werden, welche von einem Kontrollimpuls ausgelöst wird, welcher seinerseits von der muskelelektrischen Tätigkeit des Stumpfs vom Patienten abgeleitet wird. Die Einrichtung versetzt den Amputierten in den Stand, das künstliche Bein als einen Reaktionspunkt für seine übriggebliebenen Muskeln zu benutzen.

Das Aufsteigen von Treppen und Rampen mit der Prothese wird durch Gebrauch eines mechanischen Kniegelenks möglich gemacht, das, wenn gesperrt, nicht weiter gebeugt werden kann, das aber weiter frei ist, sich unter Kontrolle des Patienten zu strecken.

Die, in diesem Bericht beschriebene Prothese verwendet ein ziemlich ungewöhnliches Kontrollsystem. Impulsanstoss wird an Stelle von Impulsintegration benutzt. Eine logische Vorrichtung hält das Spürsystem zurück, sobald der Knieverschluss erfolgt ist. Das ganze System wird automatisch wieder eingestellt, sobald der Patient das künstlichen Bein völlig ausstreckt.

Ein Patient kann trainiert werden, in einer Sache von Stunden das Bein zum Steigen von Treppen zu benutzen.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Anderson, M. H., Bray, J. J. andHennessy, Ch. A. (1960)Prasthetic Principles—Above Knee Amputations. Thomas, Springfield.

    Google Scholar 

  • Anderson, M. H. (1963) Biomechanical considerations in the design of a long leg brace. InBiomedical Science Instrumentation, Vol. 1, pp. 385–394, Plenum Press, New York.

    Google Scholar 

  • Battye, C. K., Nightingale, A. andWhillis, J. (1955) The use of myoelectric currents in the operation of prostheses.J. Bone Jt Surg.37B (3), 506–510.

    Google Scholar 

  • Battye, C. K. andJoseph, K. (1966) An investigation by telemetering of the activity of some muscles in walking.Med. Electron. biol. Engng4 (2).

  • Benning, H. (1967) Bericht über eine Reise nach Sweden anlässlich der Seventh International Conference on Medical und Biological Engineering in der Zeit von 13. bis 20.8.1967.Orthop. Tech.20 (12), 333–339.

    Google Scholar 

  • Boenik, U. (1970) Einweisung in der Regnell-Hydraulik.Orthop. Tech.22 (10), 271–276.

    Google Scholar 

  • Bottomley, A. H. andCowell, T. K. (1962) An artificial hand controlled by the nerves.New Scientist21, 668–671.

    Google Scholar 

  • Bottomley, A. H., Kinnier-Wilson, A. B. andNightingale, A. (1963) Muscle substitutes and myo-electric control.J. Br. I.R.E.26 (6), 439–448.

    Google Scholar 

  • Burger, S. (1968) Die Rohrskelett-Prothese und ihre Anwendung.Orthop. Tech.20 (11), 303–305.

    Google Scholar 

  • Burglen, E. M., Traub, B. andWilson, A. B. Jr (1967) Immediate postsurgical prosthetics in the management of lower extremity amputees.Veterans Administration Report TR 10-5, April 1967, 20–25, Washington D.C., Veterans Administration.

    Google Scholar 

  • Habermann, H. (1968) Die Funktionen des Prothesenfussen speziell bei der Oberschenkelprothese.Orthop. Tech.20 (12), 339–343.

    Google Scholar 

  • Hiemen, H. (1968) Betrachtungen zur. bioelektrischen Versorgung von Oberarmamputierte.Orthop. Tech.20 (12), 348–352.

    Google Scholar 

  • Horn, G. W. (1963a) Electtronica e Automazione negli Apparecchi di Protesi Ortopedica.Scienza e Tecnica Ortopedica in Italia e all'Estero. No. 80, p. 81. Ass. Naz. Produttori Presidi Ortopedici, Treviso.

    Google Scholar 

  • Horn, G. W. (1963) Muscle voltage moves artificial hand.Electronics30 (41), 34–36.

    Google Scholar 

  • Horn, G. W. (1964) Il controllo mioelettrico dell' arto artificiale e la mano cibernetica di vercelli.Automaz. Automat.9 (3), 5–8.

    MathSciNet  Google Scholar 

  • Horn, G. W. (1966) Forearm prosthesis maintains skin contact: amplifier devised to use high-impedance.Medical Tribune, New York, March.

  • Horn, G. W. (1970) Lockable knee is key to artificial leg control.Electronics43 (19), 204–205.

    Google Scholar 

  • Inman, V. T. (1968) Conservation of energy in ambulation.Bull. prostheti. Res.10 (9), 26–36.

    Google Scholar 

  • Jacobs, K. J. (1968a) Relazione su Protesi cosmetiche di femore.Scienza e Tecnica Ortopedica in Italia e all'Estero No. 110, pp. 43–53. Ass. Naz. Produttori Presidi Ortopedici, Treviso.

    Google Scholar 

  • Jacobs, K. J. (1968b) Die Rohrskelett Prothese und ihre Anwendung.Orthop. Tech.20 (11), 293–303.

    Google Scholar 

  • Jansen, K. (1965) Amputations.Bull. prosthet. Res.10 (4), 5–40.

    Google Scholar 

  • Klopsteg, P. E. Wilson, P. D. (1954)Human Limbs and their Substitutes. McGraw-Hill, New York.

    Google Scholar 

  • Lewis, E. A. andBernstock, W. M. (1968) Clinical application study of the Henschke-Mauch hydraulik model. A swing and stance control for above-knee prostheses; design adjustment and installation.Bull. prosthet. Res.10 (10), 35–60.

    Google Scholar 

  • Lozach, Y., Lippay, A. L., Sherman, E. D. andGingras, G. (1967) Helping hands.Electronics40 (16), 125–129.

    Google Scholar 

  • Mauch, H. A. (1967a) Research and development in the field of artificial limbs.Bull prosthet. Res.10 (8), 143–169.

    Google Scholar 

  • Mauch, H. A. (1967b) Stance control for above-knee artificial legs—design considerations in the S−N−S knee.Bull. prosthet. Re.,10 (10), 61–72.

    Google Scholar 

  • Murdoch, G. (1968) Myoplastic techniques.Bull. prosthet. Res.10 (9), 4–13.

    Google Scholar 

  • Napier, J. (1967) The antiquity of human walking.Scient. Am.216 (4), 56–68.

    Article  Google Scholar 

  • Peisner, W., Antonelli, D. andWaring, W. (1968) Zener simulates muscle signals.Electronics41 (13), 111–112.

    Google Scholar 

  • Peizer, E. (1967a) VAPC report.Bull. prosthet. Res.10 (7), 199–247.

    Google Scholar 

  • Peizer, E. (1967b) VAPC report.Bull. prosthet. Res.10 (8), 208–300.

    Google Scholar 

  • Radcliffe, C. W. andLamoreoux, L. (1968) UC-BL pneumatic swing control unit for above knee prostheses —design, adjustment and installation.Bull. prosthet. Res.10 (10), 73–89.

    Google Scholar 

  • Scott, R. N. andThompson, G. B. (1963) Orthotic System Research Progress Report No. 2. University of New Brunswick,Dept. of Elect. Engng Rept. No. 63-5.

  • Scott, R. N. (1967) Myo-electric control of prostheses and orthoses.Bull. prosthet. Res.10 (7), 93–114.

    Google Scholar 

  • Tosberg, W. A. (1962)Upper and Lower Extremity Prostheses. Thomas, Springfield.

    Google Scholar 

  • Vodovnik, L. andMcLeod, W. D. (1965) Electronics detours for broken nerve paths.Electronics38 (19), 110–116.

    Google Scholar 

  • Vodovnik, L., Dimitrijevic, M. R., Preveéc, T. andLogar, M. (1966)Electronic Walking-Aid for Patients with Peroneal Palsy. Electrotechnical Faculty University of Ljubljana, Ljubljana, Jugoslavia.

    Google Scholar 

  • Vodovnik, L. (1964) An electromagnetic brake activated by eyebrow muscles.Electron. Engng. 10.

  • Zavod SRS za Rehabilitacijo Invalidov (1968)Rezis Kovalni Project Funktionalna Elektricna Stimulacija Extremided. Ljubljana, (Jugoslavia Institut za Rehaboilitacija Invalidov.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Horn, G.W. Electro-control: am EMG-controlled A/K prosthesis. Med. & biol. Engng. 10, 61–73 (1972). https://doi.org/10.1007/BF02474569

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02474569

Keywords

Navigation