Verbesserung der Nachweisempfindlichkeit der instrumentellen Neutronenaktivierungsanalyse: Anticomptonspektrometrie mit einem zentralen Ge(Li)-Bohrlochdetektor

Improvement of sensitivity in instrumental neutron activation analysis by means of anticompton spectrometry with a central well-type Ge(Li)-detector

  • Originalabhandlungen
  • Published:
Fresenius' Zeitschrift für analytische Chemie Aims and scope Submit manuscript

Abstract

To improve the sensitivity of trace element determinations in biological samples by instrumental neutron activation analysis, a modified anticompton spectrometer was constructed which operates with a well-type Ge(Li) diode as the primary detector. To absorb the compton scattered gamma rays the central detector is surrounded by two cylindrical NaI(Tl) crystals (8″ dia × 4″ ht). Each of these crystals has a 1.5″ radius semi-circular slot cut in its face to provide optimum design of the multicrystal device. The performance of the spectrometer is specified by the Y-88 figure of merit for 1 cc gamma sources: 20.4 at 122 keV, 4.0 at 662 keV, and 2.9 at 1332 keV.

The operation of the spectrometer in the coincidence mode results in a significant suppression of the bremsstrahlung background, a fact highly advantageous in the analysis of phosphorus-rich biological material. In the analysis of blood serum, for e.g., the background produced by P-32 can be reduced by a factor of 28.4 at 136 keV. Thus it was possible to improve the detection limit by a factor of 5 in the determination of selenium via the 136 keV gamma rays from Se-75 when compared with the conventional measurement at cap of a single large coaxial Ge(Li)-detector.

Zusammenfassung

Zur Steigerung der Nachweisempfindlichkeit der instrumentellen neutronenaktivierungs-analytischen Spurenelementbestimmung in biologischen Proben wurde ein Anticompton-Spektrometer aufgebaut, bei dem erstmals ein Ge(Li)-Bohrlochdetektor als Zentraldetektor eingesetzt wird. Als Fängerdetektoren für die aus dem Ge(Li) austretenden Compton-Streuquanten dienen zwei zylindrische NaJ(Tl)-Kristalle (8″ Ø × 4″ mit jeweils halbzylindrischer Bohrung), die den Zentraldetektor in optimaler Geometrie umgeben. Die Leistungsfähigkeit der Anlage für 1 ml Quellen wird durch folgende Y-88 Güteziffern demonstriert: 20,4 bei 122 keV, 4,0 bei 662 keV, 2,9 bei 1332 keV.

Im Koinzidenzbetrieb wird eine wirkungsvolle Bremsstrahlungsunterdrückung erreicht. Für Serumproben beträgt der P-32-Unterdrückungsfaktor 28,4 bei 136 keV, so daß die Nachweisgrenze für die Se-75-Bestimmung über die 136 keV-Linie um einen Faktor 5 gegenüber einer at cap-Messung der Probe mit einem 95 cm3 Ge(Li)-Vollkristall erniedrigt werden kann.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Baedecker, P. A.: Anal. Chem. 43, 405 (1971)

    Google Scholar 

  2. Behne, D., Brätter, P., Geßner, H., Hube, G., Merz, W., Rösick, U.: Fresenius Z. Anal. Chem. 278, 269 (1976)

    Google Scholar 

  3. Brätter, P., Gawlik, D., Lausch, J., Rösick, U.: Proc. 1976 Intern. Conf. Modern Trends in Activation Analysis, Vol. 1, p. 257 (1976)

    Google Scholar 

  4. Brätter, P., Lausch, J., Lausch, K., Rösick, U.: HMI-Bericht Nr. 206 (1976)

  5. Cooper, J. A.: Nucl. Instr. Meth. 82, 273 (1970)

    Google Scholar 

  6. Cooper, J. A., Perkins, R. W.: Nucl. Instr. Meth. 94, 29 (1971)

    Google Scholar 

  7. Cooper, J. A., Perkins, R. W.: Nucl. Instr. Meth. 99, 125 (1972)

    Google Scholar 

  8. Cooper, J. A., Rancitelli, L. A., Perkins, R. M.: J. Radioanal. Chem. 6, 147 (1970)

    Google Scholar 

  9. Currie, L. A.: Anal. Chem. 40, 586 (1968)

    Google Scholar 

  10. Erdtmann, G., Soyka, W.: Jül-1003-AC, 2. Aufl. (1974)

  11. Rogers, V. C.: Anal. Chem. 42, 807 (1970)

    Google Scholar 

  12. Taczanowski, S.: Nucl. Instr. Meth. 123, 409 (1975)

    Google Scholar 

  13. Verplanke, J. G.: Nucl. Instr. Meth. 96, 557 (1971)

    Google Scholar 

  14. Yule, H. P.: J. Radioanal. Chem. 15, 695 (1973)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Herrn Prof. Dr. K. E. Zimen zum 65. Geburtstag gewidmet.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rösick, U., Brätter, P. Verbesserung der Nachweisempfindlichkeit der instrumentellen Neutronenaktivierungsanalyse: Anticomptonspektrometrie mit einem zentralen Ge(Li)-Bohrlochdetektor. Z. Anal. Chem. 286, 336–342 (1977). https://doi.org/10.1007/BF00431193

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00431193

Navigation