Summary
In a previous paper [1] the sensitivity of radiation balance meters with a polyethylene dome was discussed. In these discussions it appeared that the absorption properties of the thermopile and the transmission of the polyethylene play an important role in sensitivity.
A number of polyethylene domes has been measured on their transmission power before as well as after use, to get information on the effect of use. The absorption factor for longwave thermal radiation of Parsons optical matt black has been calculated by comparing net radiation data from the Funk net radiometer and calculated data from separate measurements of incoming and outgoing radiation by the Schulze balance radiometer.
Zusammenfassung
In einer früheren Mitteilung [1] wurde die Empfindlichkeit von Strahlungsbilanzmessern mit Lupolenhalbkugel diskutiert; dabei ergab sich, daß bei der Empfindlichkeit die Absorptionseigenschaften der Thermosäule und die Durchlässigkeit des Lupolens eine bedeutende Rolle spielen.
Um den Einfluß der Benützung abzuklären, wurde bei einer Anzahl Lupolenhalbkugeln die Durchlässigkeit vor und nach erfolgter Exposition gemessen. Das Absorptionsvermögen von Parsons mattem Schwarzlack für langwellige Wärmestrahlung wurde berechnet durch eine Vergleichung von Nettostrahlungswerten nach dem Funk-Strahlungsmesser und von berechneten Werten separater Messungen der einfallenden und der ausgehenden Strahlung mit einem Schulze-Strahlungsbilanzmesser.
Résumé
Dans un précédent mémoire [1], on a discuté la sensibilité des appareils de mesure du bilan radiatif recouverts de coupoles de polyéthylène. Il en découlait que la dite sensibilité dépend avant tout des propriétés d'absorption de la pile thermo-électrique et du pouvoir de transmission de l'enveloppe de polyéthylène.
Afin de déterminer les effets découlant de leur utilisation, on a mesuré la perméabilité au rayonnement d'un certain nombre de coupoles de polyéthylène avant et après leur exposition. On a calculé le pouvoir d'absorption en rayonnement calorifique de grande longueur d'ondes de la laque noire et mate de Parsons en comparant les relevés de rayonnement net d'un radiomètre de Funk aux valeurs calculées de mesures séparées du rayonnement reçu et émis mesures faites au moyen d'un bilanmètre de rayonnement de Schulze.
Similar content being viewed by others
Abbreviations
- ia :
-
ratio\({{\tau _{sh}^{pl} } \mathord{\left/ {\vphantom {{\tau _{sh}^{pl} } {\alpha _{lo}^p }}} \right. \kern-\nulldelimiterspace} {\alpha _{lo}^p }}\)
- d :
-
ratio\({{\alpha _{sh}^p } \mathord{\left/ {\vphantom {{\alpha _{sh}^p } {\tau _{lo}^{pl} }}} \right. \kern-\nulldelimiterspace} {\tau _{lo}^{pl} }}\)
- H :
-
radiant flux densityW·m −2
- R :
-
reciprocal sensitivityW·m −2 mV −1
- t :
-
time
- α:
-
absorption factor
- ϑ:
-
transmission factor
References
Wartena, L., C. L. Palland, andA. Koetsier: Some Experiences on the Measuring of Long-wave Radiation Fluxes. Arch. Met. Geoph. Biokl. B,14, 189 (1966).
Abraham, F. F.: Determination of Long-Wave Atmospheric Radiation. J. Met.17, 291 (1960).
Ambach, W., E. Beschorner andH. Hoinkes: Über die Eichung des Strahlungsbilanzmessers nachR. Schulze (Lupolengerät). Arch. Met. Geoph. Biokl., B,13, 76 (1965).
Schulze, R.: Über die Verwendung von Polyäthylen für Strahlungsmessungen. Arch. Met. Geophy. Biokl., B,11, 211 (1962).
Marchgraber, R., andR. M. Armstrong: Solar Radiation Measurement Instrumentation USA SRDL. Technical report 2251, U. S. Army Signal Research & Development Laboratory, Fort Monmouth, N. J. (Personal communication).
Author information
Authors and Affiliations
Additional information
With 1 Figure
Rights and permissions
About this article
Cite this article
Palland, C.L., Wartena, L. Investigations on the calibration factor of the Schulze and the Funk radiation balance meters and comparison of some measured results. Arch. Met. Geoph. Biokl. B. 16, 95–104 (1968). https://doi.org/10.1007/BF02248869