In vivo measurement of lead in bone using X-ray fluorescence

, and

Published under licence by IOP Publishing Ltd
, , Citation L J Somervaille et al 1985 Phys. Med. Biol. 30 929 DOI 10.1088/0031-9155/30/9/005

0031-9155/30/9/929

Abstract

The factors affecting the accuracy and minimum detectable concentration of in vivo tibia lead measurement are discussed, and it is demonstrated that the use of a 109Cd source in a backscatter geometry and using the 88 keV coherently scattered photon for normalisation optimises both criteria. The measurement is shown to be independent of variations in source-sample distance, thickness of overlying tissue and tibia size and shape. Applying the same technique in vitro to samples of human tibia and metatarsals, it is shown that the results are not significantly different (p approximately=0.9) from atomic absorption spectrometry results from another laboratory. The results of Monte Carlo dose distribution calculations are presented and compared with measurements using thermoluminescent dosemeters: the mean absorbed dose to a 20 cm leg section is less than 0.1 mGy (10 mrad) and the maximum absorbed skin dose is 0.45 mGy (45 mrad). For this dose the minimum detectable lead concentration is 10 mu g g-1. Finally, the technique has been applied to groups of normals and occupationally exposed workers, and the means have been shown to be significantly different, namely 10 and 31 mu g g-1 respectively. In the normal subjects tibia lead correlated strongly with age (r=0.63, p<0.001).

Export citation and abstract BibTeX RIS

10.1088/0031-9155/30/9/005