Practical approach to electron beam dosimetry at extended SSD

, , and

Published under licence by IOP Publishing Ltd
, , Citation Joanna Cygler et al 1997 Phys. Med. Biol. 42 1505 DOI 10.1088/0031-9155/42/8/003

0031-9155/42/8/1505

Abstract

This paper presents a practical approach to clinical dosimetry for electron beams at extended source-to-surface distance (SSD). Characteristics of electron beams from a Siemens MD-2 accelerator are presented at a nominal SSD of 100 cm and an extended SSD of 115 cm. Relative output factors are measured at both 100 and 115 cm SSDs for a range of square field sizes for each beam energy. The change of output with SSD does not follow the inverse square law, if the nominal SSD is used. This deviation is larger for lower beam energy and smaller field sizes. An effective SSD, , has been determined at for each beam energy as a function of field size. A comprehensive approach to treatment time calculations is proposed.

Export citation and abstract BibTeX RIS