Dosimetry with tissue-equivalent ionisation chambers in fast neutron fields for biomedical applications

and

Published under licence by IOP Publishing Ltd
, , Citation J Zoetelief and J J Broerse 1983 Phys. Med. Biol. 28 503 DOI 10.1088/0031-9155/28/5/005

0031-9155/28/5/503

Abstract

The use of calibrated issue-equivalent (TE) ionisation chambers is commonly considered to be the most practical method for total absorbed dose determinations in mixed neutron-photon fields for biomedical applications. The total absorbed dose can be derived from the charge produced within the cavity of an ionisation chamber employing a number of physical parameters. To arrive at the charge produced in the cavity several correction factors have to be introduced which are related to the operational characteristics of the chambers. Information on the operational characteristics of four TE ionisation chambers is presented in relation to the ion collection, density and composition of gas in the cavity, wall thickness and effective point of measurement. In addition, some recent results from an ionisation chamber operated at high gas pressures (up to 8 MPa(80 bar)) are presented. The total chamber operated at high gas pressures (up to 8 MPa(80 bar)) are presented. The total absorbed doses derived from TE ionisation chambers show agreement within the uncertainty limits with results from other independent dosimetry methods, i.e., differential fluence measurements and a TE calorimeter. Conscientious experimentation and a common data base can provide dosimetry results with TE ionisation chambers with variations of less than +or-2%.

Export citation and abstract BibTeX RIS