Perturbation correction for ionometric determination of absorbed dose in a graphite phantom for 60Co gamma rays

Published under licence by IOP Publishing Ltd
, , Citation M Boutillon 1983 Phys. Med. Biol. 28 375 DOI 10.1088/0031-9155/28/4/006

0031-9155/28/4/375

Abstract

Calculations have been made of the perturbation correction to be applied to a flat cavity ionisation chamber. This makes possible the use of such a chamber as a primary standard of absorbed dose in graphite for 60Co gamma rays. The good agreement between ionometric measurements and the calorimetric measurements of national laboratories of metrology gives strong support to the reliability of the calculated correction.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/28/4/006