TECHNICAL NOTE

Measurement of the voltage of diagnostic X-ray tubes using K-characteristic fluorescence

and

Published under licence by IOP Publishing Ltd
, , Citation M Davison and J Law 1972 Phys. Med. Biol. 17 572 DOI 10.1088/0031-9155/17/4/010

0031-9155/17/4/572

Abstract

Reports results obtained by both the ratio method and the single detector method, using X-rays generated both at constant potential and from a self-rectified unit. The agreement between the two methods is excellent for X-rays from both types of generator, indicating that the assumption in the ratio method is valid when the precautions discussed in this note are taken.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/17/4/010