Comparison of graphite-to-water absorbed-dose transfers for Co photon beams using ionometry and Fricke dosimetry

, and

Published under licence by IOP Publishing Ltd
, , Citation B Chauvenet et al 1997 Phys. Med. Biol. 42 2053 DOI 10.1088/0031-9155/42/11/004

0031-9155/42/11/2053

Abstract

To derive the absorbed dose to water from a standard of absorbed dose to graphite, the metrology laboratories which apply such a method usually make use of cavity ionization chambers as transfer instruments. In addition, the BNM-LPRI has tested, as such instruments, two types of Fricke dosimeter in its cobalt-60 beam. The two procedures are compared and their results are found to be in good agreement (the difference is less than 0.1%). Both procedures are then taken into account for the calculation of the reference value of absorbed dose to water.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/42/11/004