Traceable calibration of ultraviolet meters used with broadband, extended sources

, and

Published under licence by IOP Publishing Ltd
, , Citation A J Coleman et al 2000 Phys. Med. Biol. 45 185 DOI 10.1088/0031-9155/45/1/313

0031-9155/45/1/185

Abstract

A calibration system has been developed to provide increased accuracy in the measurement of the irradiance responsivity appropriate for UV meters used with broadband, extended sources of the type employed in phototherapy. The single wavelength responsivity of the test meter is obtained in the wavelength range 250-400 nm by intercomparison with a transfer standard meter in a narrow, monochromatic beam. Traceability to primary standard irradiance scales is provided via the National Measurement System with a best uncertainty of 7% (at 95% confidence). The effective responsivity of the test meter, when used with broadband extended sources, is calculated using the measured spectral and angular response of the meter and tabulated data on the spectral and spatial characteristics of the source radiance. The uncertainty in the effective responsivity, independent of the source variability, is estimated to be 10% (at 95% confidence). The advantages of this calibration system over existing approaches are discussed.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/45/1/313