Synchrotron measurements of the energy response functions of CdTe Medipix2 MXR detectors with pixel pitches of 110 and 165 μm

, , , , , and

Published 13 December 2012 Published under licence by IOP Publishing Ltd
, , Citation M Zuber et al 2012 JINST 7 C12018 DOI 10.1088/1748-0221/7/12/C12018

1748-0221/7/12/C12018

Abstract

The energy resolution of photon counting pixel detectors depends strongly on the pixel pitch. Small pixels increase the effects of characteristic X-rays and the sharing of charges between adjacent pixels. To investigate the effects of different pixel pitches, measurements with Medipix2 MXR detectors attached to CdTe sensors with pixel pitches of 110 and 165 μm were carried out at the Topo-Tomo beamline of the ANKA synchrotron at the Karlsruhe Institute of Technology (KIT). Photo peaks are measured at a number of energies and are used to calculate the fraction of events in the photo peak compared to the total. Finally, we investigate the linearity of the energy calibration curves as a function of the pixel pitch.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/1748-0221/7/12/C12018