Abstract
How a DSC result is influenced by the particle size distribution of a powder sample is shown, and a simple and optimal method to be included in a routine DSC analysis (e.g., purity determination) to improve the reliability of the analysis is proposed. In case ofα-Al2O3 powder, most reliable heat capacity data can be obtained by preparing a powder with a self-similar particle size distribution with a distribution constant of 0.7, and by compressing it under a pressure of 1.5 MPa for a duration of 5 min or longer.
Similar content being viewed by others
References
Purity Determinations by Thermal Methods, ASTM Special Technical Publication 838, R. L. Blaine and C. K. Schoff (eds.), 1984.
J. Chiu and P. G. Fair, Thermochim. Acta, 34 (1979) 267.
W. W. Hillstrom, Thermal Conductivity, 16 (1979) 483.
T. J. Coutts, Thin Solid Films, 38 (1976) 313.
D. S. MacLachlan, Solid State Commun., 60 (1986) 21.
See, for example, R. Ozao and M. Ochiai, Thermochim. Acta, 198 (1992) 289.
Y. Nishimoto, H. Ichimura and Y. Teramoto, Netsu-sokutei, 17 (1990) 61.
R. Ozao and M. Ochiai, J. Thermal Anal., 40 (1993) 1331.
D. A. Ditmars and T. B. Douglas, J. Res. Nat. Bur. Stand. A, 75A (1971) 401.
D. A. Ditmars et al., J. Res. Nat. Bur. Stand., 87 (1982) 159.
See, for example, G. Deutscher, A. Kaptulnik and M. Rappaport, in Percolation Structures and Processes, Ch.10, G. Deutscher, R. Zallen and J. Adler (Eds.), Adam Hilger, Bristol 1983.
N. D. Osborne, Part. Syst. Charact., 9 (1992) 202.
Author information
Authors and Affiliations
Additional information
This work is partially supported by the Research Fund of North Shore College of SONY Institute. R.O. wishes to express her gratitude to the support.
Rights and permissions
About this article
Cite this article
Ozao, R., Ogura, H., Ochiai, M. et al. Thermal analysis of powdered alumina materials. Journal of Thermal Analysis 49, 961–970 (1997). https://doi.org/10.1007/BF01996782
Issue Date:
DOI: https://doi.org/10.1007/BF01996782