Abstract
The evaluation of accuracy and efficiency of the frequency-modulated simultaneous Atomic Absorption Spectrometry (FremsAAS) has been extended to an arrangement with EDL as light sources. Fundamental calibrations have been worked out for As, Se and Sb using a graphite furnace as well as hydride generation in combination with a heated quartz tube as atomization unit. The characteristic data are in good agreement with results obtained by conventional single-channel AAS instruments. Determinations in three standard reference materials with different complex matrices resulted in complete agreement with the certified values.
Similar content being viewed by others
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Edel, H., Erber, D., Lehnert, R. et al. Frequency-modulated simultaneous Atomic Absorption Spectrometry (FremsAAS): Determination of As, Se and Sb. Fresenius J Anal Chem 355, 292–294 (1996). https://doi.org/10.1007/s0021663550292
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s0021663550292