Summary A device is described which will prevent the destruction of a thermal conductivity detector in a gas chromatograph, for instance if there is a serious error in the operation of the latter. The device is coupled with the power supply of the detector; it will switch off the power supply whenever the errors occur. Experimental results of the study are given, together with recommendations for the use of the circuit-breaker.