ISSN:
1573-4994
Keywords:
Monochromator
;
calibration
;
mercury
;
argon
;
near-infrared
Source:
Springer Online Journal Archives 1860-2000
Topics:
Physics
Notes:
Abstract The emission spectra of four mercury vapor lamps are used to obtain wavelength calibration curves for the double-grating emission monochromator of a spectrofluorimeter. The use of second- and third-order diffraction lines and emission lines from the argon carrier gas provides a rich spectrum, which extends well into the near-infrared spectral region and produces an improved calibration curve. More than 60 emission lines are listed between 250 and 900 nm, which is sufficient to produce an extremely accurate monochromator calibration. Additional second- and third-order lines can be used to even longer wavelengths (〉1200 nm). The effectiveness of three scattering surfaces is compared.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1023/A:1009414126732