    Weinheim : Wiley-Blackwell
    Journal of High Resolution Chromatography 16 (1993), S. 663-665 
    ISSN: 0935-6304
    Keywords: Micro-LC ; Amperometric detector ; Nitrophenols ; Chemistry ; Analytical Chemistry and Spectroscopy
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Chemistry and Pharmacology
    Notes: An amperometric-type electrochemical detector is proposed for use in capillary HPLC. Its cell (volume 60 nl) contains a cylindrical working electrode for detecting the presence of electroactive compounds. The noise level is 3-5 pA depending on the potential. The minimum detectable concentrations of the o-, m-, and p- isomers of nitrophenol (after separation of a 20 μl sample on a fused silica column) lies between 4 and 18 ng/l. The relative standard deviation of the height of the chromatographic peaks of phenols present at concentrations of 10 ppb does not exceed 0. 08 (n = 5).
    Additional Material: 5 Ill.
    Type of Medium: Electronic Resource