Publication Date:
2019-07-13
Description:
Potentials were measured using a beam of soft X-rays in air at 2 x 10 to the -5 power Torr. Ions were detected by a continuous-dynode electron multiplier after they passed through a retarding field. Ultimate resolution depends upon the diameter of the X-ray beam which was 3 mm. When the fields in the region of interest were such to disperse the ions, only a small fraction were detected and the method of measurement was not very reliable. Yet reasonable data could be collected if the ions traveled in parallel paths toward the detector. Development should concentrate on increasing the aperture of the detector from the pinhole which was used to something measured in centimeters. Also increasing the strength of the source would provide a stronger signal and more reliable data. Measurements were made at an estimated ion current to 10 to the -15 power A from a 10 cm length of the X-ray beam, this current being several orders of magnitude below what would have a perturbing effect on the region to be measured. Consequently, the source strength can be increased and prospects for this method of measurement are good.