Publication Date:
2019-07-13
Description:
We report on the performance of our NTD-Ge microcalorimeters. To date, the spectral resolution for x-ray and gamma-ray lines from radioactive sources and laboratory plasmas is 4.8 eV in the entire 1 - 6 keV band and 52 eV at 60 keV. Technical details responsible for this performance are presented as well as an innovative electro-thermal approach for enhancing count-rate capability.
Keywords:
Instrumentation and Photography
Type:
Ninth International Workshop on Low Temperature Detectors; Jul 22, 2001 - Jul 27, 2001; Madison, WI; United States
Format:
text
Permalink