Description:
A technique for improving the measurement certainty with the BB series (Smith et al., 1972) of electrically calibrated calorimeters used in high-energy lasers is described. The technique is based on monitoring the energy which is backscattered from the meter and monitoring the overspill radiation impinging on the calorimeter at the entrance aperture. The design and performance of a second generation BB meter is discussed and compared to that of the original device in terms of number of electrical calibrations, the residual standard deviation of electrical calibration, the calibration constant for laser energy, the correcting factor for systematics, inaccuracy, imprecision, and uncertainty.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
PB-285319
,
IEEE Transactions on Instrumentation and Measurement; IM-27; Mar. 197