Publikationsdatum:
2019-06-28
Beschreibung:
An optical witness plate program was implemented to monitor buildup of molecular contamination in the clean room during the assembly and testing of the Halogen Occulation Experiment (HALOE) instrument. Travel plates to monitor molecular contamination when the instrument is not in the clean room are also measured. The instrument technique is high-resolution transmission spectroscopy in the 3 micron spectral region using a Fourier transform spectrometer. Witness specimens of low index of refraction, infrared transmitting material are used for contaminant monitoring and for spectral signature analysis. Spectral signatures of possible molecular contamination are presented. No condensible volatile material contamination of HALOE optical witness specimens have yet been found.
Schlagwort(e):
INSTRUMENTATION AND PHOTOGRAPHY
Materialart:
NASA-TM-4081
,
L-16500
,
NAS 1.15:4081
Format:
application/pdf
Permalink