A technique has been developed to dynamically calibrate fine wire thermocouples of 0.025 mm diam and smaller to 1625 K in less than 10 sec. In this method, the varying physical parameters present under typical dynamic heating conditions are simulated, and therefore, wire deterioration can be related to time, temperature, pressure, and atmosphere. A description of the equipment and procedures is given along with the preliminary results that correlate well with thermoelectric theory.