Publication Date:
2019-07-13
Description:
The dual differential radiometer (DDR) was tested to determine its capability for measuring suspended sediment and chlorophyll in turbid waters. Measurements were obtained from a boat dock and from a helicopter with combinations of sample and reference filters with peak transmissions at various wavelengths. Water samples were taken concurrently and were analyzed for light scattering, particle count, and total chlorophyll. Least-squares estimates of the linear relationship between DDR output and the water parameters yielded correlation coefficients of less than 0.7. Under the turbid water conditions of the present tests, the DDR did not accurately measure either suspended sediment or chlorophyll. A precise knowledge of the spectral signatures of various pollutants might enable appropriate filters to be selected for tuning the DDR to monitor a particular pollutant.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
Remote sensing of earth resources. Volume 4; Mar 24, 1975 - Mar 26, 1975; Tullahoma, TN
Format:
text
Permalink