ALBERT

All Library Books, journals and Electronic Records Telegrafenberg

feed icon rss

Your email was sent successfully. Check your inbox.

An error occurred while sending the email. Please try again.

Proceed reservation?

Export
  • 1
    Electronic Resource
    Electronic Resource
    Springer
    Microchimica acta 126 (1997), S. 263-266 
    ISSN: 1436-5073
    Keywords: ICP-MS ; thorium ; uranium ; urine ; body-fluid ; trace elements
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Abstract ICP-MS (inductively coupled plasma mass spectrometry) is shown as a very sensitive method for quantitative determination of Th and U concentration and excretion analysis in urine without any sample pretreatment. The current standard method for incorporation monitoring applies alpha-spectrometry, a very tedious and time consuming technique. ICP-MS offers an attractive alternative for monitoring of thorium and uranium body burdens in occupationally exposed subjects and also larger groups of the general population. A limit of determination of 0.5 ng/L in aqueous solutions and 1 ng/L in urine samples for both elements was achieved, with a precision of about ± 10% in the concentration range of appr. 10 ng/L. Due to the lack of a suitable reference material, the accuracy of the method was tested by comparing some of the results with those obtained by α-spectrometry, especially for U. There was a sufficient agreement on both results.
    Type of Medium: Electronic Resource
    Location Call Number Expected Availability
    BibTip Others were also interested in ...
Close ⊗
This website uses cookies and the analysis tool Matomo. More information can be found here...