ISSN:
1436-5073
Keywords:
luminol chemiluminescence
;
HPLC
;
reaction detector
;
amino acid
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Notes:
Abstract A detector for HPLC is based on suppression of chemiluminescence from the Cu(II)-luminol-peroxide reaction. Analytes which complex Cu(II) reduce the free Cu(II) concentration and thus the observed chemiluminescence intensity. The degree of chemiluminescence suppression is a measure of the analyte concentrations. Detection limits are in the range of 1 pmole-1 nmole for amino acids (both primary and secondary without derivatization), CN−, amines, catecholamines, catechol, and aminoglycoside antibiotics. The detection approach is demonstrated with an ion-exchange separation of amino acids, a reverse phase separation of catecholamines, and an ion-pair separation of the three components of gentamicin C in commercial gentamicin sulfate.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF01201720