    ISSN: 1612-1112
    Keywords: Column liquid chromatography ; Microcapillary column ; Gasoline group analysis ; Conventional IR detection
    Source: Springer Online Journal Archives 1860-2000
    Topics: Chemistry and Pharmacology
    Notes: Summary Slurry-packed microcapillary column liquid chromatography with conventional infrared detection is described for the group-type analysis of gasolines. The results indicate that the three eluted groups, saturates, olefins and aromatics, from the microcapillary column are easily detected by conventional microcomputerized infrared spectrometer coupled with a Teflon flow-cell with a good signal-to-noise ratio. This study shows the feasibility of microcapillary column liquid chromatography combined with a low-cost, conventional infrared spectrometer as a routine tool which allows the analyst to use an LC-IR system effectively and economically.