0049-8246
Keywords:
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Physics
Notes:
The widely used formalism for quantitative XRF analysis based on the correction of characteristic line intensity for absorption effects is shown to be inadequate when thick samples with low total mass attenuation coefficients are analysed (biological samples, solutions, etc.). The geometry factor is not a constant function of the collimator's shape and dimensions, but it is found that in an analysed matrix it depends on the line energy and penetration depth. An experimental and a theoretical model for the determination of the geometrical correction are proposed and tested on a set of samples. The concentrations obtained are in excellent agreement with the values expected, whereas the model based on the correction of intensities only for absorption effects gives an error of at least 25% for thick samples.
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/xrs.1300160307