ISSN:
0935-6304
Keywords:
GC instrument conversion
;
All-glass GC-2 system
;
Phenols
;
Hydrocarbons
;
Terpenoids
;
Linear response
;
Reproducibility
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Two regular GC instruments were modified with injection port liners, split-mode injection systems, pressure regulators, effluent traps and needle valves to produce an all-glass GC-2 system. This simple and inexpensive system allows the quantitative analyses of phenols, hydrocarbons, acids, sterols, terpenoids, and alkaloids with excellent reproducibility and linear response. Even more impressive was its ability to analyze a C45-terpenoid alcohol, solanesol, as its TMS-derivative (M.W. 702).
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240030104
Permalink