ISSN:
1618-2650
Keywords:
Best. von Phosphor in Wasser
;
Spektralphotometrie, AutoAnalyzer, Meerwasser, partikulärer Phosphor, Mo-Komplex
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
Description / Table of Contents:
Summary After filtration of sea water through glass fiber filters, the filters with the residue are digested by heating with sulphuric acid and hydrogen peroxide. The following determination of phosphate is based on the formation of a phosphomolybdenum complex. It has been modified in such a way that the phosphorus content of the samples can be determined with a continuous-flow system with the TechniconAutoAnalyzer direct after digestion. Two methods are described for different sensitivities of 0.1–10μmol P · l−1.
Notes:
Zusammenfassung Der Rückstand bei Filtration von Meerwasser durch Glasfaserfilter wird mit Schwefelsäure und Wasserstoffperoxid aufgeschlossen. Die anschließende Phosphatbestimmung basiert auf der Bildung eines Phosphormolybdän-Komplexes. Sie wurde modifiziert, so daß der Phosphor-Gehalt der Proben sofort nach Aufschluß mit dem Durchflußsystem des Technicon-AutoAnalyzers bestimmt werden kann. Zwei Methoden mit unterschiedlicher Empfindlichkeit (von 0,1–10μmol P · l−1) werden beschrieben.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF00481264
Permalink