ISSN:
0935-6304
Keywords:
Gas chromatography-mass spectrometry
;
Capillary columns
;
Stationary phases
;
Silicones
;
Thermal degradation
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
The applicability of capillary columns for gas chromatography is often limited by stationary phase degradation at elevated temperatures. In order to achieve a better understanding of column thermostability, column bleed products have been analyzed qualitatively by gas chromatography-mass spectrometry (GC-MS). Three silicone stationary phases were studied: SE-52, SE-54, and OV-1701. Each of these was immobilized in the columns. The proportion of cyclics containing diphenyl in column bleed from SE-52 reflected to a large extent the composition of the polymer, while for SE-54 the proportion of such cyclics was unexpectedly high. The polar moiety of OV-1701, the cyanopropyl(phenyl)siloxy unit, was found to exert a highly destabilizing effect on the polymer, and the thermal degradation products consisted mainly of cyclosiloxanes containing this unit.