ISSN:
1573-8906
Source:
Springer Online Journal Archives 1860-2000
Topics:
Electrical Engineering, Measurement and Control Technology
Notes:
Conclusions Investigations showed that the TVB-6 transducer can be used with standard measuring equipment (ÉPP-09, numerical voltmeters, etc.) in measuring absolute pressures of 1.33–1330 Pa. The relative mean square error of the transducer is 5%. Its service life is 100 h. The transducer, which is made commercially, is compact and cheap. However, it requires a highly stable feed current and its characteristic is nonlinear.