ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
In this article we describe an electrochemical experimental setup for cyclic voltammetry operating from low (1 V/s) to very high (200 kV/s) scan rates. The system is designed to achieve an acceptable cost and to be user friendly for nonexperts. The instrumentation is based on a commercial arbitrary wave form generator, a digital oscilloscope, a general purpose interface bus interface to a PC, and a laboratory-built potentiostat. There are no complicated manual operations, instead the software creates dialog with an operator to control commands for the instruments. Requirements for the wave form synthesis and the data acquisition are given. Various data-smoothing methods for treating 8-bit fast scan data are evaluated. The performance was tested on the oxidation of ferrocene in acetonitrile. The estimated value of the standard heterogeneous rate constant, k0=0.8±0.1 cm s−1, is in reasonable agreement with the most recent reports from other laboratories. © 2000 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1150540