Measurements of the mean static free-stream gas density have been made in two Langley Research Center helium facilities, the 3-inch leg of the high-Reynolds-number helium complex and the 22-inch hypersonic helium tunnel. Rayleigh scattering of a CW argon ion laser beam at 514.5 nm provided the basic physical mechanism. The behavior of the scattered signal was linear, confirmed by a preliminary laboratory study. That study also revealed the need to introduce baffles to reduce stray light. A relatively simple optical system and associated photon-counting electronics were utilized to obtain data for densities from 10 to the 23rd to 10 to the 25th per cu m. The major purpose, to confirm the applicability of this technique in the hypersonic helium flow, was accomplished.
Keywords:
INSTRUMENTATION AND PHOTOGRAPHY
Type:
AIAA PAPER 79-1086
,
American Institute of Aeronautics and Astronautics, Thermophysics Conference; Jun 04, 1979 - Jun 06, 1979; Orlando, FL