Schlagwort(e):
Gas chromatography, capillary
;
sampling, sample discrimination, split ratio
;
many factors influencing real amount sampled
;
external calibration to be replaced by inner standard technique
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Quelle:
Wiley InterScience Backfile Collection 1832-2000
Thema:
Chemie und Pharmazie
Notizen:
The effect of pressure changes in the vaporizing GC injectors on the split ratio is investigated. It is shown that the true split ratio depends on a number of parameters and may strongly differ from the split ratio expected from the flows set before injection. Quantitation with split injection based on external standards has to be carried out very carefully (or should be replaced by the method with internal standard). The split ratio may strongly change during the splitting of the sample. There are a number of mechanisms causing some sample components to be retarded and hence to be split by a different ratio. The split injector is a device which still requires further development.