    Surface and Interface Analysis 14 (1989), S. 595-597 
    ISSN: 0142-2421
    Keywords: Chemistry ; Polymer and Materials Science
    Source: Wiley InterScience Backfile Collection 1832-2000
    Topics: Physics
    Notes: A new type of sputtered neutral mass spectrometer, named SNART (Sputtered Neutral Analysis - Riken Type), was evaluated. SNART uses an argon plasma that is driven by an electron beam. The plasma is used for both sputtering of the sample and post-ionizing the sputtered neutral atoms. A depth profile of multilayered Ge/Si was taken by SNART to evaluate depth resolution. The quantitativeness of SNART was also checked using a stainless-steel sample. The results were compared with those by conventional SIMS and glow discharge optical emission spectrometry. The following features of SNART were confirmed: (1) high post-ionization efficiency; (2) high depth resolution; (3) high sputtering rate; (4) no matrix effect on post-ionization efficiency; and (5) high quantitativeness.
    Additional Material: 5 Ill.