ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
Charge exchange spectroscopy is one of the key ion diagnostics on the DIII-D tokamak. It allows measurement of impurity densities, toroidal and poloidal rotation speeds, ion temperatures, and the radial electric field. For the 2000 experimental campaign, we have replaced the intensified photodiode array detectors on the edge portion of the system with advanced charge-coupled device (CCD) detectors mounted on faster (f/4.7) Czerny–Turner spectrometers equipped with toroidal mirrors. The combination has improved the photoelectron signal level by about a factor of 20 and the signal to noise by a factor of 2–8, depending on the absolute signal level and readout mode. A major portion of the signal level improvement comes from the improved quantum efficiency of the back-illuminated, thinned CCD detector (70% to 85% quantum efficiency for the CCD versus 10% for the image intensifier) with the remainder coming from the faster spectrometer. The CCD camera also allows shorter minimum integration times: 0.33 ms while archiving to computer memory and 0.15 ms using temporary storage on the CCD chip. © 2001 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1319363