Publication Date:
2019-07-13
Description:
The Digicon has been chosen as the detector for the High Resolution Spectrograph and the Faint Object Spectrograph of the Space Telescope. Both tubes are 512 channel, parallel-output devices and feature CsTe photocathodes on MgF2 faceplates. Using a computer-assisted test facility, the tubes have been characterized with respect to diode array performance, photocathode response (1100-9000 A), and imaging capability. Data are presented on diode dark current and capacitance distributions, pulse height resolution, photocathode quantum efficiency, uniformity and blemishes, dark count rate, distortion, resolution, and crosstalk.
Keywords:
SPACECRAFT INSTRUMENTATION
Type:
Seminar on Advances in focal plane technology; Feb 04, 1980 - Feb 05, 1980; Los Angeles, CA