Summary A simple and efficient tantalum strip atomizer is described suitable for use in the atomic absorption determination of trace elements in water. The cell operates with an inexpensive Powerstat-transformer unit to provide the necessary electrothermal heat for atomization. It can be easily fitted in the place of the burner in atomic absorption instruments. The atomizer has been used to determine Ag, Cd, Pb, and Zn in tap water samples and Zn in the presence of albumin.
Notes:
Zusammenfassung Die elektrische Erhitzung des Tantalstreifens wird mit Hilfe einer preisgünstigen Energiequelle (Powerstat/Transformator) vorgenommen, die leicht anstelle des Brenners in den üblichen Geräten untergebracht werden kann. Das Verfahren wurde zur Bestimmung von Ag, Cd, Pb und Zn in Leitungswasser sowie von Zink neben Albumin angewendet.