ISSN:
0935-6304
Keywords:
GC×GC
;
comprehensive two-dimensional GC
;
thermal modulation
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
---A jet of cool gas is used to locally cool a section of modulator tube in the presence of the stirred oven bath of a GC×GC instrument. Local cooling decouples the temperature of the modulator tube from that of the first dimension column, which was 100 meters long. Overall resolution of the GC×GC experiment was improved as a result. Another consequence of the jet-cooled thermal modulation structure is the elimination of moving parts in the GC oven. By pulsing cold and hot jets of gas onto a modulator tube with solenoid valves, two stage thermal modulation can be obtained without the complexity of moving parts in the vicinity of the capillary columns.
Additional Material:
1 Ill.
Type of Medium:
Electronic Resource