A new technique called dynamic measurement access (DMA) is presented in this article which was developed for relatively easy, fast, and accurate thermal conductivity measurements. Utilizing a commercial cryocooler adapted for this experiment, the DMA technique allows the measurement of typically 400 data points in a temperature range between 300 and 10 K, during an elapsed time of approximately 90 min. This technique provides very good precision and requires a very simple experimental setup, which is described in this article. © 2001 American Institute of Physics.