ISSN:
0884-3996
Keywords:
Diabetes
;
albuminuria
;
chemiluminescent immunoassay
;
acridinium ester
;
Chemistry
;
Biochemistry and Biotechnology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Biology
,
Chemistry and Pharmacology
Notes:
A simple chemiluminescent immunoassay (CLIA) for urinary albumin has been developed based on the use of a chemiluminescent acridinium ester-labelled human albumin and a commercially available antiserum. It includes two incubation steps and a second polyethylene glycol-assisted antibody separation. The sensitivity of detection is 0.016 mg/l, the assay working range is 0.1-5 mg/l, and the inter-assay CVs are ≤ 15%. Using 10- and 50-fold sample dilutions in assay buffer, a wide working range (1-250 mg/l) is obtained covering normal and pathological conditions. Timed overnight urine samples (bed rest conditions) were collected on three consecutive days for each patient. Albumin excretion rate (AER) was 4.7 ± 2.7 μg/min (x ± SD), range 1-15.9 μg/min in 36 healthy subjects (17♂, 19♀, ages 4-56 years), with day-to-day variations of 28.5 ± 20% (x ± SD), range 3.3-76.1%. The use of an acridinium ester as a chemiluminescent (CL) label overcomes the disadvantages of short shelf-life and health and safety hazards associated with radioisotopes. Results compare favourably with those obtained using a commercially available RIA kit.
Additional Material:
3 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/bio.1170030403