ISSN:
1588-2780
Source:
Springer Online Journal Archives 1860-2000
Topics:
Chemistry and Pharmacology
,
Energy, Environment Protection, Nuclear Power Engineering
Notes:
Abstract A new Compton suppression system was established. A low energy HPGe detector (LO-AX) is used as the analyzing detector. It is completely shielded by a 50%n-type HPGe and two NaI(TI) detectors. Experiments show that a substantial Compton suppression effect is also achieved in the low energy region down to 15 keV. The capability of applying this system in INAA was evaluated for 13 elements (As, Cd, Co, Cr, Hg, Mo, Ni, Sb, Se, Sr, Th, U and Zn) in biological and environmental samples. The benefits and problems of using this Compton suppression spectrometer in INAA are examined for each element.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02034462