ISSN:
0935-6304
Keywords:
Gas chromatography
;
Capillary, fused silica
;
Priority pollutants
;
Volatile organics
;
Extractable organics
;
Pesticides
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
Acrolein, acrylonitrile, volatile organics, base/neutral extractables, acid extractables, and pesticide fraction organics are analyzed in a single chromatographic run. This is accomplished by injecting the extractable organics, then applying a cryogenic focus to the front of the column, and desorbing the volatile components. The column utilized for this procedure is a fused silica open tubular (FSOT) column coated with the liquid phase SE-54. Minimum detection limits of this procedure by electron impact GC/MS are shown to be≤ 1 μg/liter in tests conducted on actual water samples.
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1002/jhrc.1240040403