A completely implantable epidural pressure telemetry system designed for accurate measurement of intracranial pressure (ICP) is described. The implant device is batteryless, providing unlimited operating life. The described system uses a capacitive pressure transducer with excellent long-term stability. Once detected with the transducer and converted to a frequency with the oscillator electronics, the pressure signal is digitized. It is then telemetered without the possibility of further degradation. After detection with the small external module, the data can be retransmitted by a radio link for complete patient mobility or the energizer signal pickup module can be wired to a bedside readout unit. Continuous data are available from the system so that the dynamic ICP changes reflecting arterial blood pressure can be observed and used for diagnosis.
Keywords:
MAN/SYSTEM TECHNOLOGY AND LIFE SUPPORT
Type:
Biotelemetry and Patient Monitoring; 5; 2, 19; 1978