ISSN:
0006-3592
Keywords:
green fluorescent protein
;
sensor
;
on-line monitoring
;
quantitation
;
Chemistry
;
Biochemistry and Biotechnology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Biology
,
Process Engineering, Biotechnology, Nutrition Technology
Notes:
We present an intensity based sensor designed for on-line monitoring of green fluorescent protein, a revolutionary marker of protein expression. The device consisted of a blue light emitting diode as the excitation source. A band pass excitation filter cut off light longer than 490 nm. The light was directed into a bifurcated optical fiber bundle with the common end inserted into a stainless steel housing equipped with a quartz window. The fiber bundle and stainless steel housing are steam sterilizable. The emission radiation was collected through a long wave pass filter to reject the excitation light shorter than 505 nm and was detected by a photomultiplier tube. The signal was amplified and sent to a computer for recording time course data. The sensor was tested in an Escherichia coli fermentation of JM105 transformed with pBAD-GFP. The on-line signal was compared to off-line fluorescence spectrophotometer measurements. The on-line profile closely followed the off-line. Western blot data showed that with a time shift, the sensor was able to both continuously and quantitatively monitor expression of green fluorescent protein on-line in real time. © 1997 John Wiley & Sons, Inc. Biotechnol Bioeng 55:921-926, 1997.
Additional Material:
6 Ill.
Type of Medium:
Electronic Resource
Permalink