ISSN:
1741-0444
Keywords:
Correlation
;
Foetal heart rate
;
Modulus difference
;
Ultrasonic foetal monitoring
Source:
Springer Online Journal Archives 1860-2000
Topics:
Biology
,
Chemistry and Pharmacology
,
Medicine
Notes:
Abstract Microprocessor-based correlation processing is now widely used for the estimation of foetal heart rate from ultrasonic Doppler signals. The paper describes a technique having a, similar performance, to correlation but which will run three times faster on a microprocessor without an onboard hardware multiplier. This speed improvement results from replacing each multiplication operation of the correlation algorithm with a subtraction and a modulus operation. The algorithm, is thus referred to as the modulus difference algorithm. Typical waveforms are presented illustrating the responses of the two processors to continuous wave Doppler foetal heart signals for both wide (50 Hz) and narrow (4 Hz) preprocessor bandwidths. A quantitative comparison is given of the performance of the two methods in estimating the period of a synthetic signal with additive white Gaussian noise. This simulation represents the wideband preprocessor case. A theoretical derivation of the outputs for the narrowband preprocessor case is given, and actual outputs are shown for comparison. A tutorial approach is adopted throughout the paper.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1007/BF02443924