Electron microscopy
;
EELS
;
feedback system
;
peak stabilization
;
Life and Medical Sciences
;
Cell & Developmental Biology
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Natural Sciences in General
Notes:
The position of the zero-loss peak in electron energy-loss spectra was sensed with a double photo diode. A signal, proportional to the disturbances from a central position, was amplified and fed back into a deflection coil in order to compensate for the origin of the disturbances. Thus, slow variations of the position of characteristic edges in the EEL spectrum could be reduced by a factor 100, and 60 Hz oscillations could be reduced by a factor 5.