    Journal of clinical monitoring and computing 9 (1992), S. 179-182 
    ISSN: 1573-2614
    Keywords: computer ; control ; model ; pharmacokinetic ; propofol ; target
    Source: Springer Online Journal Archives 1860-2000
    Topics: Computer Science , Medicine
    Notes: Abstract A portable target controlled infusion system for propofol has been developed based on a Psion hand-held microcomputer and the Ohmeda 9000 syringe pump. The system uses a pharmacokinetic model which describes the distribution and elimination of propofol to achieve and maintain any selected target blood concentration. Target blood concentrations of 1 μg/ml, 3 μg/ml and 5 μg/ml were selected in laboratory trials and the cumulative volumes delivered by the Psion system each minute were compared with the theoretical output calculated by the pharmacokinetic model. The results obtained showed that the computer system delivered volumes which were always within 2% of the theoretical values. This system offers a convenient and simple method of maintaining anaesthesia using propofol.