ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
,
Electrical Engineering, Measurement and Control Technology
Notes:
A new spherical grating monochromator with translational and rotational assembly including a normal incidence mount (SGM-TRAIN) is under construction at BL 5A of the UVSOR Facility, IMS, and its design concept is reported. The following points have been taken into account in the design: (1) Energy range of 5–250 eV, (2) use both of undulator and bending magnet radiation, (3) linear and circular polarizations, (4) spectral purity, (5) resolution, and (6) length of beamline. It should be stressed that the SGM-TRAIN is useful for experiments where circularly polarized synchrotron radiation is used. Moreover, the SGM-TRAIN has the advantage that the small emittance of storage ring and the large space for a long beamline are not necessary. © 1995 American Institute of Physics.
Type of Medium:
Electronic Resource
URL:
http://dx.doi.org/10.1063/1.1146512
Permalink