Gas chromatography/mass spectrometry
;
Capillary columns, fused silica
;
Response factors, interlaboratory
;
Chemistry
;
Analytical Chemistry and Spectroscopy
Source:
Wiley InterScience Backfile Collection 1832-2000
Topics:
Chemistry and Pharmacology
Notes:
A quality control procedure for Fused Silica Capillary Column Gas Chromatography/Mass Spectroscopy (FSCC GC/MS) was tested and shown to produce similar response factors at five Spectroscopy (FSCC GC/MS) was tested and shown to produce similar response factors at five laboratories. The average relative standard deviation (RSD) of interlaboratory response factors for fifty three compounds of environmental interest determined at each laboratory at nominal injected weights of 20, 100, and 200 ng was found to be 18.9% for four of the five laboratories which participated in this work. These data demonstrate that similar response factors can be attained in routine GC/MS measurements when adequate quality control is maintained.