An instrument has been developed and tested for the continuous measurement of volatile organic compounds (VOC) in air. The system consists of a gas chromatograph equipped with a dedicated sampling device that allows the sample to be transferred to a cooled microtrap via sampling loops (10, 100, 250 ml) or via a direct pump transfer to the trap. The microtrap is placed in the chromatographic oven just below a modified split-splitless injector, allowing direct liquid injection for calibration of the system; the injector is in communication with the sampling valve equipped with the loop and the sampling pump. The system allows 24-hour sampling and analysis of a large number of VOC (up to 25 individual hydrocarbons ranging from C2—C9) and also polar volatile organic compounds PVOC. Thanks to the particular trap geometry, a minimum consumption of liquid nitrogen (between 150—300 ml) is needed for each analytical run and no water managing system is normally required for humid air samples.